An official FDA notice concerns Systems units containing affected software sql beckman coulter chemistry analyzer au480 ref UDI-di code: b96692 14987666542965 b96693. Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results. Published 2026-10-07.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

492

Recall number

Z-3332-2026

Official notice

2026-10-07

The hazard

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 14987666542965, UDI-DI 14987666542972, UDI-DI 14987666540671

Event Ids

99588

Recall Numbers

Z-3332-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.

Units affected

492

Hazard

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

BG · Bulgaria

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CN · China

Affected area 8

CR · Costa Rica

Affected area 9

CY · Cyprus

Affected area 10

CZ · Czech Republic

Affected area 11

DE · Germany

Affected area 12

EG · Egypt

Affected area 13

FR · France

Affected area 14

GT · Guatemala

Affected area 15

HR · Croatia

Affected area 16

HU · Hungary

Affected area 17

IE · Ireland

Affected area 18

IN · India

Affected area 19

IT · Italy

Affected area 20

JP · Japan

Affected area 21

KR · South Korea

Affected area 22

LV · Latvia

Affected area 23

MY · Malaysia

Affected area 24

PK · Pakistan

Affected area 25

PL · Poland

Affected area 26

RO · Romania

Affected area 27

RS · Serbia

Affected area 28

SA · Saudi Arabia

Affected area 29

TR · Turkey

Affected area 30

TW · Taiwan

Affected area 31

ZA · South Africa

Company & contacts

Manufacturer

Beckman Coulter Mishima K.K.

Phone

Not provided

Brand: Beckman Coulter Mishima K.k. Contact: Beckman Coulter Mishima K.K..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3332-2026

Event ID

99588

Units affected

492

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button