Open official notice

Electrolase(r) hyfrecator(r) electrodes - sharp, disposable hyfrecator(r) tips, ref catalog no. 7-100-8cs, sterile, Rx only -- conmed

Official notice

30 Jan 2013

RecallScope risk estimate:MediumFDA

ELECTROLASE(R) HYFRECATOR(R) ELECTRODES - SHARP, Disposable Hyfrecator(R) Tips, REF/Catalog No. 7-100-8CS, Sterile, Rx ONLY -- CONMED CORPORATION Electrosurgical, Cutting and Coagulation and Accessories.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

313,190

Recall number

Z-0718-2013

Official notice

2013-01-30

The hazard

ConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury. The compromised seals were discovered prior to use.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

120628X

Event Ids

63981

Recall Numbers

Z-0718-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ELECTROLASE(R) HYFRECATOR(R) ELECTRODES - SHARP, Disposable Hyfrecator(R) Tips, REF/Catalog No. 7-100-8CS, Sterile, Rx ONLY -- CONMED CORPORATION Electrosurgical, Cutting and Coagulation and Accessories.

Units affected

313,190

Hazard

ConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury. The compromised seals were discovered prior to use.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CO · Colombia

Affected area 4

CZ · Czech Republic

Affected area 5

GT · Guatemala

Affected area 6

HK · Hong Kong

Affected area 7

IL · Israel

Affected area 8

IN · India

Affected area 9

IT · Italy

Affected area 10

JP · Japan

Affected area 11

KW · Kuwait

Affected area 12

MX · Mexico

Affected area 13

MY · Malaysia

Affected area 14

NO · Norway

Affected area 15

NZ · New Zealand

Affected area 16

PK · Pakistan

Affected area 17

PT · Portugal

Affected area 18

SA · Saudi Arabia

Affected area 19

TH · Thailand

Affected area 20

UY · Uruguay

Company & contacts

Brand

Manufacturer

ConMed Corporation

Phone

Not provided

Brand: Conmed. Contact: ConMed Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0718-2013

Event ID

63981

Units affected

313,190

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 30 Jan 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button