Open official notice

Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source

Official notice

9 Dec 2015

RecallScope risk estimate:MediumFDA

Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

9

Recall number

Z-0359-2016

Official notice

2015-12-09

The hazard

Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

72586

Recall Numbers

Z-0359-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.

Units affected

9

Hazard

Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · TX

Affected area 1

US · United States

Affected area 2

MX · Mexico

Company & contacts

Brand

Manufacturer

ConMed Corporation

Phone

Not provided

Brand: Conmed. Contact: ConMed Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0359-2016

Event ID

72586

Units affected

9

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Dec 2015. The next one won’t wait for you to search.