Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended

Official notice

2 Oct 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended. The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle. Published 2013-10-02.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

266

Recall number

Z-2263-2013

Official notice

2013-10-02

The hazard

The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

66247

Recall Numbers

Z-2263-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended use of the REPS is washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical devices.

Units affected

266

Hazard

The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Affected area 4

ES · Spain

Company & contacts

Brand

Manufacturer

Steris Corporation

Phone

Not provided

Brand: Steris. Contact: Steris Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2263-2013

Event ID

66247

Units affected

266

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Oct 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button