Open official notice

Steris Lighting and Visualization Systems, Sterile Light Handle Cover (LB53)- Surgical light handle covers, including disposable surgical

Official notice

10 May 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Steris Lighting and Visualization Systems, Sterile Light Handle Cover (LB53)- Surgical light handle covers, including disposable surgical. Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient Published 2023-05-10.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,358,520

Recall number

Z-1475-2023

Official notice

2023-05-10

The hazard

Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

9604187

Event Ids

92040

Recall Numbers

Z-1475-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Steris Lighting and Visualization Systems, Sterile Light Handle Cover (LB53)- Surgical light handle covers, including disposable surgical light handle covers are designed to fit over surgical light handles for sterile surgeon control. REF: LB53

Units affected

1,358,520

Hazard

Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CA · Canada

Company & contacts

Brand

Manufacturer

Steris Corporation

Phone

Not provided

Brand: Steris. Contact: Steris Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1475-2023

Event ID

92040

Units affected

1,358,520

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 May 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button