Open official notice

Infinion cx lead kit, 50cm, model: sc-2317-50, material: m365sc2317500; infinion cx lead kit, 70cm, model: sc-2317-70, material

Official notice

12 Aug 2026

Last checked

22 Sept 2026

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Infinion cx lead kit, 50cm, model: sc-2317-50, material: m365sc2317500; infinion cx lead kit, 70cm, model: sc-2317-70, material. Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement. Published 2026-06-17.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

1,079

Recall number

Z-2879-2026

Official notice

2026-06-17

The hazard

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

98974

Recall Numbers

Z-2879-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC2317700. part of the Boston Scientific Spinal Cord Stimulator (SCS) Systems.

Units affected

1,079

Hazard

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · NY

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CO · Colombia

Affected area 7

CZ · Czech Republic

Affected area 8

DE · Germany

Affected area 9

ES · Spain

Affected area 10

FR · France

Affected area 11

GB · United Kingdom

Affected area 12

GR · Greece

Affected area 13

HU · Hungary

Affected area 14

IE · Ireland

Affected area 15

IL · Israel

Affected area 16

IT · Italy

Affected area 17

NL · Netherlands

Affected area 18

PT · Portugal

Affected area 19

RS · Serbia

Company & contacts

Manufacturer

Boston Scientific Neuromodulation Corporation

Phone

Not provided

Brand: Boston Scientific Neuromodulation. Contact: Boston Scientific Neuromodulation Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class I

Recall details

Recall number

Z-2879-2026

Event ID

98974

Units affected

1,079

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 Aug 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button