Image not available in RecallScope.
Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These
Official notice
25 Mar 2020
Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, Rx Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electrocautery Probe, J Tip, 5MM 0250070443-PKG, StrykeFlow Electrocautery Probe, L Tip, 5MM 0250070444-PKG, StrykeFlow Electrocautery Probe, Ball Tip, 5MM 0250070445-PKG, StrykeFlow Electrocautery Probe, Needle Tip 5MM 0250070446-PKG, StrykeFlow Electrocautery Probe, Spoon Tip, 5MM
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
525
Recall number
Z-1508-2020
Official notice
2020-03-25
The hazard
The hazard
Potential diff. In the length of the sheath tube and the sheath/probe. A diff. In length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Company & contacts
Company & contacts
Brand: Stryker. Contact: Stryker Corporation.
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.