Open official notice

The 3.0MM Neuro Match Head Soft Touch bur (5820-107-030) is intended to be used to cut bone and bone cement during neurosurgery, including

Official notice

29 Jan 2025

RecallScope risk estimate:MediumFDA

The 3.0MM Neuro Match Head Soft Touch bur (5820-107-030) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/ Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. The 3.0MM Prec Neuro Match Head bur (5820-107-530) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. These burs are sold as single use devices without their own IFU.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

491

Recall number

Z-0980-2025

Official notice

2025-01-29

The hazard

There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-030 but package contains a 5820-107-530 device).

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

21341057

Event Ids

95996

Recall Numbers

Z-0980-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

The 3.0MM Neuro Match Head Soft Touch bur (5820-107-030) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/ Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. The 3.0MM Prec Neuro Match Head bur (5820-107-530) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. These burs are sold as single use devices without their own IFU.

Units affected

491

Hazard

There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-030 but package contains a 5820-107-530 device).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CO · Colombia

Affected area 3

GB · United Kingdom

Affected area 4

NL · Netherlands

Affected area 5

SE · Sweden

Company & contacts

Brand

Manufacturer

Stryker Corporation

Phone

Not provided

Brand: Stryker. Contact: Stryker Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0980-2025

Event ID

95996

Units affected

491

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Jan 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button