Open official notice

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes

Official notice

15 Jul 2026

Last checked

22 Sept 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns OmniTract Surgical Retractor System, which includes 633 of the firm s product codes. The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used. Published 2026-06-05.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

633

Recall number

Z-2714-2026

Official notice

2026-06-05

The hazard

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

99193

Recall Numbers

Z-2714-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

Units affected

633

Hazard

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

BG · Bulgaria

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CL · Chile

Affected area 10

CN · China

Affected area 11

CZ · Czech Republic

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

EE · Estonia

Affected area 15

EG · Egypt

Affected area 16

ES · Spain

Affected area 17

FI · Finland

Affected area 18

FR · France

Affected area 19

GB · United Kingdom

Affected area 20

GR · Greece

Affected area 21

HK · Hong Kong

Affected area 22

HU · Hungary

Affected area 23

IE · Ireland

Affected area 24

IL · Israel

Affected area 25

IN · India

Affected area 26

IS · Iceland

Affected area 27

IT · Italy

Affected area 28

JO · Jordan

Affected area 29

JP · Japan

Affected area 30

KR · South Korea

Affected area 31

KW · Kuwait

Affected area 32

LT · Lithuania

Affected area 33

MT · Malta

Affected area 34

MX · Mexico

Affected area 35

MY · Malaysia

Affected area 36

NL · Netherlands

Affected area 37

NO · Norway

Affected area 38

NZ · New Zealand

Affected area 39

PK · Pakistan

Affected area 40

PL · Poland

Affected area 41

PT · Portugal

Affected area 42

SA · Saudi Arabia

Affected area 43

SE · Sweden

Affected area 44

SG · Singapore

Affected area 45

SI · Slovenia

Affected area 46

TH · Thailand

Affected area 47

TR · Turkey

Affected area 48

TW · Taiwan

Affected area 49

ZA · South Africa

Company & contacts

Manufacturer

Integra LifeSciences Corp. (NeuroSciences)

Phone

Not provided

Brand: Integra Lifesciences. Contact: Integra LifeSciences Corp. (NeuroSciences).

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2714-2026

Event ID

99193

Units affected

633

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button