Open official notice

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Official notice

18 Dec 2019

RecallScope risk estimate:MediumFDA

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

820

Recall number

Z-0613-2020

Official notice

2019-12-18

The hazard

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

84216

Recall Numbers

Z-0613-2020

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

C7000 - CUSA¿ Clarity Ultrasonic Tissue Ablation System

Units affected

820

Hazard

The 36 kHz Sterilization Tray (C7136 Rev C / 72904491 Rev A) system does not meet the Dry Time Criteria for a 30-minute drying cycle. The current CUSA Clarity Operator s manual states 30 minutes for the minimum Dry Time for the 36 kHz Sterilization Tray and 36 kHz Components but 40 minutes is required for complete drying.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · AR

Location 3

US · United States · AZ

Location 4

US · United States · CA

Location 5

US · United States · CO

Location 6

US · United States · CT

Location 7

US · United States · FL

Location 8

US · United States · GA

Location 9

US · United States · IA

Location 10

US · United States · ID

Location 11

US · United States · IL

Location 12

US · United States · IN

Location 13

US · United States · KS

Location 14

US · United States · KY

Location 15

US · United States · LA

Location 16

US · United States · MA

Location 17

US · United States · MD

Location 18

US · United States · MI

Location 19

US · United States · MN

Location 20

US · United States · MO

Location 21

US · United States · MS

Location 22

US · United States · NC

Location 23

US · United States · NE

Location 24

US · United States · NJ

Location 25

US · United States · NY

Location 26

US · United States · OH

Location 27

US · United States · OK

Location 28

US · United States · OR

Location 29

US · United States · PA

Location 30

US · United States · SC

Location 31

US · United States · TN

Location 32

US · United States · UT

Location 33

US · United States · VA

Location 34

US · United States · WA

Location 35

US · United States · WI

Location 36

US · United States · WV

Affected area 1

US · United States

Company & contacts

Manufacturer

Integra LifeSciences Corp.

Phone

Not provided

Brand: Integra Lifesciences. Contact: Integra LifeSciences Corp..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0613-2020

Event ID

84216

Units affected

820

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 18 Dec 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button