Open official notice

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Official notice

29 Jul 2026

Last checked

22 Sept 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android). For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia. Published 2026-06-16.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

3,465

Recall number

Z-2778-2026

Official notice

2026-06-16

The hazard

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

99228

Recall Numbers

Z-2778-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Units affected

3,465

Hazard

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

CH · Switzerland

Affected area 4

DE · Germany

Affected area 5

DK · Denmark

Affected area 6

ES · Spain

Affected area 7

FI · Finland

Affected area 8

IS · Iceland

Affected area 9

NO · Norway

Affected area 10

SE · Sweden

Company & contacts

Manufacturer

Medtronic MiniMed, Inc.

Phone

Not provided

Brand: Medtronic Minimed. Contact: Medtronic MiniMed, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2778-2026

Event ID

99228

Units affected

3,465

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.