An official FDA notice concerns Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13. When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. For mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction. Published 2026-10-07.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

6

Recall number

Z-3307-2026

Official notice

2026-10-07

The hazard

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. For mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00884838125698, UDI-DI 00884838128422

Event Ids

99834

Recall Numbers

Z-3307-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Ingenia Blue Seal SE and Blue Seal XE; Model number (REF): 1. 782185 (SE), 2. 782192 (XE); Software Versions: R11, R12 and R13

Units affected

6

Hazard

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. For mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CL · Chile

Affected area 10

CN · China

Affected area 11

CO · Colombia

Affected area 12

CY · Cyprus

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

DK · Denmark

Affected area 16

EG · Egypt

Affected area 17

ES · Spain

Affected area 18

FI · Finland

Affected area 19

FR · France

Affected area 20

GB · United Kingdom

Affected area 21

GR · Greece

Affected area 22

HK · Hong Kong

Affected area 23

HU · Hungary

Affected area 24

IE · Ireland

Affected area 25

IL · Israel

Affected area 26

IN · India

Affected area 27

IT · Italy

Affected area 28

JO · Jordan

Affected area 29

JP · Japan

Affected area 30

KR · South Korea

Affected area 31

LU · Luxembourg

Affected area 32

LV · Latvia

Affected area 33

MT · Malta

Affected area 34

MX · Mexico

Affected area 35

MY · Malaysia

Affected area 36

NL · Netherlands

Affected area 37

NO · Norway

Affected area 38

NZ · New Zealand

Affected area 39

PE · Peru

Affected area 40

PK · Pakistan

Affected area 41

PL · Poland

Affected area 42

PT · Portugal

Affected area 43

RO · Romania

Affected area 44

RS · Serbia

Affected area 45

SA · Saudi Arabia

Affected area 46

SE · Sweden

Affected area 47

SG · Singapore

Affected area 48

SK · Slovakia

Affected area 49

TH · Thailand

Affected area 50

TW · Taiwan

Affected area 51

UY · Uruguay

Affected area 52

VN · Vietnam

Affected area 53

ZA · South Africa

Company & contacts

Manufacturer

Philips North America

Phone

Not provided

Brand: Philips North. Contact: Philips North America.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3307-2026

Event ID

99834

Units affected

6

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button