Guardian 4 glucose sensor, ref: mmt-7040a, mmt-7040c1, mmt-7040c2, mmt-7040c3, mmt-7040d1, mmt-7040d2, mmt-7040ma, mmt-7040qc1

Official notice

14 Feb 2024

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Guardian 4 glucose sensor, ref: mmt-7040a, mmt-7040c1, mmt-7040c2, mmt-7040c3, mmt-7040d1, mmt-7040d2, mmt-7040ma, mmt-7040qc1. Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot. Published 2024-02-14.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

114,758

Recall number

Z-1005-2024

Official notice

2024-02-14

The hazard

Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

93675

Recall Numbers

Z-1005-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3

Units affected

114,758

Hazard

Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 ¿m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

BE · Belgium

Affected area 4

CH · Switzerland

Affected area 5

CZ · Czech Republic

Affected area 6

DE · Germany

Affected area 7

DK · Denmark

Affected area 8

ES · Spain

Affected area 9

FI · Finland

Affected area 10

FR · France

Affected area 11

GB · United Kingdom

Affected area 12

GR · Greece

Affected area 13

IE · Ireland

Affected area 14

IL · Israel

Affected area 15

IN · India

Affected area 16

IS · Iceland

Affected area 17

IT · Italy

Affected area 18

KW · Kuwait

Affected area 19

LI · Liechtenstein

Affected area 20

LU · Luxembourg

Affected area 21

NL · Netherlands

Affected area 22

NO · Norway

Affected area 23

PL · Poland

Affected area 24

PT · Portugal

Affected area 25

RO · Romania

Affected area 26

SA · Saudi Arabia

Affected area 27

SE · Sweden

Affected area 28

SI · Slovenia

Company & contacts

Manufacturer

Medtronic MiniMed, Inc.

Phone

Not provided

Brand: MINI, Medtronic Minimed. Contact: Medtronic MiniMed, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1005-2024

Event ID

93675

Units affected

114,758

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Feb 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button