Volista standop surgical light, ard568811961 vlt600df aim stp; ard568821961 vlt600df aim stp ql; ard568811911 vlt600sf aim stp

Official notice

25 Jul 2018

RecallScope risk estimate:HighFDA

An official FDA notice concerns Volista standop surgical light, ard568811961 vlt600df aim stp; ard568821961 vlt600df aim stp ql; ard568811911 vlt600sf aim stp. Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use. Published 2018-07-25.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

151

Recall number

Z-2502-2018

Official notice

2018-07-25

The hazard

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

80452

Recall Numbers

Z-2502-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.

Units affected

151

Hazard

Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CO · Colombia

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

DK · Denmark

Affected area 16

EE · Estonia

Affected area 17

EG · Egypt

Affected area 18

ES · Spain

Affected area 19

FI · Finland

Affected area 20

FR · France

Affected area 21

GB · United Kingdom

Affected area 22

HK · Hong Kong

Affected area 23

HR · Croatia

Affected area 24

HU · Hungary

Affected area 25

IE · Ireland

Affected area 26

IN · India

Affected area 27

IT · Italy

Affected area 28

JO · Jordan

Affected area 29

JP · Japan

Affected area 30

KR · South Korea

Affected area 31

LU · Luxembourg

Affected area 32

LV · Latvia

Affected area 33

MX · Mexico

Affected area 34

MY · Malaysia

Affected area 35

NL · Netherlands

Affected area 36

PK · Pakistan

Affected area 37

PL · Poland

Affected area 38

PT · Portugal

Affected area 39

RO · Romania

Affected area 40

RS · Serbia

Affected area 41

RU · Russia

Affected area 42

SA · Saudi Arabia

Affected area 43

SE · Sweden

Affected area 44

SG · Singapore

Affected area 45

SI · Slovenia

Affected area 46

SK · Slovakia

Affected area 47

TH · Thailand

Affected area 48

TR · Turkey

Affected area 49

TW · Taiwan

Affected area 50

ZA · South Africa

Company & contacts

Manufacturer

Maquet SAS

Phone

Not provided

Brand: Maquet Sas. Contact: Maquet SAS.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2502-2018

Event ID

80452

Units affected

151

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 25 Jul 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button