VNS Therapy¿ AspireHC¿ Generator, VNS Therapy¿ AspireSR¿ Generator. Model No. 105. Product Usage: United States Indications for Use

Official notice

23 Aug 2017

RecallScope risk estimate:MediumFDA

VNS Therapy¿ AspireHC¿ Generator, VNS Therapy¿ AspireSR¿ Generator. Model No. 105. Product Usage: United States Indications for Use: Epilepsy (PMA 970003): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in adults and adolescents over 12 years of age with partial onset seizures that are refractory to antiepileptic medications. Depression (PMA 970003/S050): VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments. Indications outside the U.S.: Epilepsy (Non-US): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients whose epileptic disorder is dominated by partial seizure (with or without secondary generalization) or generalized seizures that are refractory to antiepileptic medications. AspireSR (Seizure Response) features the Automatic Stimulation Mode which is intended for patients who experience seizures that are associated with cardiac rhythm increases known as ictal tachycardia. Depression (Non-US): The VNS Therapy System is indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment-intolerant depressive episode.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

396

Recall number

Z-3019-2017

Official notice

2017-08-23

The hazard

Manufacturing process used to assemble the circuit board may result in some devices experiencing a faster than expected reduction in device longevity.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

77698

Recall Numbers

Z-3019-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VNS Therapy¿ AspireHC¿ Generator, VNS Therapy¿ AspireSR¿ Generator. Model No. 105. Product Usage: United States Indications for Use: Epilepsy (PMA 970003): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in adults and adolescents over 12 years of age with partial onset seizures that are refractory to antiepileptic medications. Depression (PMA 970003/S050): VNS Therapy is indicated for the adjunctive long-term treatment of chronic or recurrent depression for patients 18 years of age or older who are experiencing a major depressive episode and have not had an adequate response to four or more adequate antidepressant treatments. Indications outside the U.S.: Epilepsy (Non-US): The VNS Therapy System is indicated for use as an adjunctive therapy in reducing the frequency of seizures in patients whose epileptic disorder is dominated by partial seizure (with or without secondary generalization) or generalized seizures that are refractory to antiepileptic medications. AspireSR (Seizure Response) features the Automatic Stimulation Mode which is intended for patients who experience seizures that are associated with cardiac rhythm increases known as ictal tachycardia. Depression (Non-US): The VNS Therapy System is indicated for the treatment of chronic or recurrent depression in patients who are experiencing a treatment-resistant or treatment-intolerant depressive episode.

Units affected

396

Hazard

Manufacturing process used to assemble the circuit board may result in some devices experiencing a faster than expected reduction in device longevity.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

BE · Belgium

Affected area 6

CA · Canada

Affected area 7

CZ · Czech Republic

Affected area 8

DE · Germany

Affected area 9

FI · Finland

Affected area 10

GB · United Kingdom

Affected area 11

HR · Croatia

Affected area 12

IE · Ireland

Affected area 13

IL · Israel

Affected area 14

IT · Italy

Affected area 15

JP · Japan

Affected area 16

NL · Netherlands

Affected area 17

NO · Norway

Affected area 18

PL · Poland

Affected area 19

SA · Saudi Arabia

Company & contacts

Manufacturer

Cyberonics, Inc

Phone

Not provided

Brand: Cyberonics. Contact: Cyberonics, Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3019-2017

Event ID

77698

Units affected

396

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Aug 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button