Open official notice

VNS Therapy Programmer, Model 3000, v1.0 System

Official notice

27 Jun 2018

RecallScope risk estimate:CriticalFDA

VNS Therapy Programmer, Model 3000, v1.0 System

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

252

Recall number

Z-2255-2018

Official notice

2018-06-27

The hazard

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

80062

Recall Numbers

Z-2255-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VNS Therapy Programmer, Model 3000, v1.0 System

Units affected

252

Hazard

Certain Model 3000 programming events can result in miscalculation of parameters stored in the Models 103, 104, 105, and 106 generators. During these programming events, the miscalculations can lead to: " Delivery of more stimulation than intended, resulting in painful stimulation or other common side effects (Model 106 only); " No stimulation in the case of device disablement (Burst Watchdog Timeout), resulting in no therapy to the patient (Model 106 only); " Delivery of less stimulation than intended, resulting in therapeutic settings not being achieved within device specification (Models 103, 104, 105, or 106); and/or " Delays or absence of the 75% and 50% battery life indicators displayed by the programming software (Models 103, 104, 105, or 106).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AL

Location 2

US · United States · CA

Location 3

US · United States · CO

Location 4

US · United States · DE

Location 5

US · United States · FL

Location 6

US · United States · GA

Location 7

US · United States · ID

Location 8

US · United States · IL

Location 9

US · United States · IN

Location 10

US · United States · ME

Location 11

US · United States · MO

Location 12

US · United States · MS

Location 13

US · United States · NC

Location 14

US · United States · NJ

Location 15

US · United States · NY

Location 16

US · United States · PA

Location 17

US · United States · TN

Location 18

US · United States · TX

Location 19

US · United States · UT

Location 20

US · United States · WA

Location 21

US · United States · WI

Affected area 1

US · United States

Company & contacts

Manufacturer

Cyberonics, Inc

Phone

Not provided

Brand: Cyberonics. Contact: Cyberonics, Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2255-2018

Event ID

80062

Units affected

252

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Jun 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.