No image published for this recall

No image published for this recall

Stryker instruments system 8 and 9 saws

Official notice

24 Apr 2025

Last checked

24 Apr 2025

Medium riskHealth Canada

An official Health Canada notice concerns Stryker instruments system 8 and 9 saws. There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025 Published 2025-04-24.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77182

Official notice

2025-04-24

The hazard

There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77182

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - System 8 Reciprocating Saw, Lot or serial number - 2409624823 1802305513, Model or catalog number - 8206-000-000

Product 2

Affected products - System 9 Sternum Saw, Lot or serial number - Not applicable., Model or catalog number - 9207-000-000

Product 3

Affected products - System 9 Reciprocating Saw, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - 9206-000-000

Risk / hazard

There is a potential for the compression spring (0038-419-000) inside the saw's drivetrain assembly to disengage during blade loading and unloading. A disengaged spring disables the blade collar from springing back to its original home position, potentially leading to blade retention failure when loading a blade. Recall Start Date: March 14, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Stryker Instruments A Division Of Stryker, Stryker Instruments System 8 And 9 Saws.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

77182

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 24 Apr 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.