Open official notice

Sara plus

Official notice

4 Apr 2022

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Sara plus. Since Nov, 2020, Arjo has become aware of 2 customer complaints on smoke and flames coming out of the lift. In both events the extent of flames was limited to the lift and no health consequences were reported. The product (devices and spare parts shipped between Oct 2016 - Jan 2022) in the market are recognized as potentially affected with the risk of cover ignition. The investigation has revealed that the failure is attributed to the insufficient design of Sara Plus PCB that is prone to damage during use. Recall start date: Mar 28, 2022 Published 2022-04-04.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64046

Official notice

2022-04-04

The hazard

Since Nov, 2020, Arjo has become aware of 2 customer complaints on smoke and flames coming out of the lift. In both events the extent of flames was limited to the lift and no health consequences were reported. The product (devices and spare parts shipped between Oct 2016 - Jan 2022) in the market are recognized as potentially affected with the risk of cover ignition. The investigation has revealed that the failure is attributed to the insufficient design of Sara Plus PCB that is prone to damage during use. Recall start date: Mar 28, 2022

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

64046

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected Products - Sara Plus, Lot or Serial Number - More than 10 numbers, contact manufacturer., Model or catalog number - HEP1001-US, HEP0001-US

Hazard

Since Nov, 2020, Arjo has become aware of 2 customer complaints on smoke and flames coming out of the lift. In both events the extent of flames was limited to the lift and no health consequences were reported. The product (devices and spare parts shipped between Oct 2016 - Jan 2022) in the market are recognized as potentially affected with the risk of cover ignition. The investigation has revealed that the failure is attributed to the insufficient design of Sara Plus PCB that is prone to damage during use. Recall start date: Mar 28, 2022

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Arjohuntleigh, Sara Plus.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

64046

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Apr 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button