Open official notice

REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Official notice

20 Mar 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument. The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes. Published 2013-03-20.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

58

Recall number

Z-0925-2013

Official notice

2013-03-20

The hazard

The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

64354

Recall Numbers

Z-0925-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument

Units affected

58

Hazard

The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

CL · Chile

Affected area 4

JP · Japan

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0925-2013

Event ID

64354

Units affected

58

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Mar 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button