Open official notice

REF 154200 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size A3 ARCOM UHMWPE CO-CR-MO Alloy Only for use in Lateral

Official notice

17 Jul 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns REF 154200 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size A3 ARCOM UHMWPE CO-CR-MO Alloy Only for use in Lateral. Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension. Published 2013-07-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

26

Recall number

Z-1695-2013

Official notice

2013-07-17

The hazard

Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

65459

Recall Numbers

Z-1695-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

REF 154200 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size A3 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing tibial component for use in the lateral compartment of the knee. The Oxford Fixed Lateral bearings are indicated for use in the lateral compartment and intended to be implemented with bone cement.

Units affected

26

Hazard

Biomet investigation found the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · IN

Location 2

US · United States · NC

Location 3

US · United States · OH

Location 4

US · United States · PA

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

NL · Netherlands

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1695-2013

Event ID

65459

Units affected

26

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jul 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.