Open official notice

Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. Prosthesis, knee

Official notice

6 Apr 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. Prosthesis, knee. The anterior/posterior (AP) dimension may be oversized by 0.016 . If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary. Published 2016-04-06.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

890

Recall number

Z-1265-2016

Official notice

2016-04-06

The hazard

The anterior/posterior (AP) dimension may be oversized by 0.016 . If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

195929

Event Ids

73400

Recall Numbers

Z-1265-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Vanguard XP Knee System Porous Femoral Component various sizes and orientation a total knee replacement system. prosthesis, knee, patella/femorotibial, semi-constrained, uncemented, porous, coated, polymer/metal/polymer

Units affected

890

Hazard

The anterior/posterior (AP) dimension may be oversized by 0.016 . If the femoral component does not fit as tightly as intended, bony ingrowth may not occur causing loosening of the femoral component. A revision surgery may be necessary.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · UT

Affected area 1

US · United States

Affected area 2

NL · Netherlands

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1265-2016

Event ID

73400

Units affected

890

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 6 Apr 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button