Percept pc brainsense deep brain stimulation implantable neurostimulator

Official notice

29 Oct 2021

Last checked

29 Oct 2021

Medium riskHealth Canada

An official Health Canada notice concerns Percept pc brainsense deep brain stimulation implantable neurostimulator. Medtronic has determined that Cardioversion may damage the electronics in the Percept PC INS device, making the INS unresponsive and non-functional, meaning that the INS cannot be turned back on. To restore stimulation therapy, surgical replacement of the INS will be required. Cessation of therapy will likely cause a return of disease symptoms. In some cases, symptoms may return with an intensity greater than was experienced prior to implant (rebound effect). In rare cases, this can constitute a medical emergency. Recall date: 2021-10-14 Published 2021-10-29.

What to do now

Immediately stop using the product and contact manufacturer.

Recall number

63534

Official notice

2021-10-29

The hazard

Medtronic has determined that Cardioversion may damage the electronics in the Percept PC INS device, making the INS unresponsive and non-functional, meaning that the INS cannot be turned back on. To restore stimulation therapy, surgical replacement of the INS will be required. Cessation of therapy will likely cause a return of disease symptoms. In some cases, symptoms may return with an intensity greater than was experienced prior to implant (rebound effect). In rare cases, this can constitute a medical emergency. Recall date: 2021-10-14

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63534

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - PERCEPT PC BRAINSENSE DEEP BRAIN STIMULATION IMPLANTABLE NEUROSTIMULATOR, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalog number - B35200

Risk / hazard

Medtronic has determined that Cardioversion may damage the electronics in the Percept PC INS device, making the INS unresponsive and non-functional, meaning that the INS cannot be turned back on. To restore stimulation therapy, surgical replacement of the INS will be required. Cessation of therapy will likely cause a return of disease symptoms. In some cases, symptoms may return with an intensity greater than was experienced prior to implant (rebound effect). In rare cases, this can constitute a medical emergency. Recall date: 2021-10-14

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Medtronic, Percept Pc Brainsense Deep Brain Stimulation Implantable Neurostimulator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

63534

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
No image published for this recall

Harmony™ delivery catheter system

Medium riskHealth Canada4 Jun 2026
CA
No image published for this recall

Octopus nuvo, octopus evo as and octopus 4

High riskHealth Canada13 May 2026
CA
No image published for this recall

Medtronic dlp retrograde cardioplegia cannulae

Medium riskHealth Canada30 Apr 2026
CA
No image published for this recall

Synchromed clinician programmer application

Critical riskHealth Canada15 Apr 2026
CA
No image published for this recall

Dlp select 3d arterial cannula and dlp elongated one piece arterial cannulae

Medium riskHealth Canada3 Jul 2026
CA
No image published for this recall

Medtronic restore™ clinician programmer application

Medium riskHealth Canada28 Nov 2025
CA
No image published for this recall

Aurora ev-icd™ mri surescan™

Medium riskHealth Canada6 Nov 2025
CA
No image published for this recall

Vanta sequentia lt clinician programmer application

Medium riskHealth Canada10 Dec 2025
CA
No image published for this recall

Carelink smartsync™ applications

Critical riskHealth Canada25 Jun 2025
CA
No image published for this recall

Inceptiv

Critical riskHealth Canada10 Jun 2025
CA
No image published for this recall

Dlp cardiac vents & sumps

Medium riskHealth Canada24 Sept 2025
CA
No image published for this recall

Dlp™ elongated one piece arterial cannulae

Medium riskHealth Canada19 Dec 2024

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 29 Oct 2021. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.