Vanta sequentia lt clinician programmer application

Official notice

10 Dec 2025

Last checked

10 Dec 2025

Medium riskHealth Canada
Watch Medtronic

An official Health Canada notice concerns Vanta sequentia lt clinician programmer application. The Vanta A71200 CP app V2.0.2465 and V2.0.2683 will display a "system error" if the user selects the start usage button twice when interrogating a new implantable neurostimulator (INS). This error prompts the user to restart the CP app and contact Medtronic for assistance. Upon subsequent interrogation of the INS, the Vanta A71200 CP app will display a "validation error" prompting the user to resolve the error in order to proceed with the INS interrogation. If the validation error is not resolved, the user is unable to complete the first attempt at interrogating the INS. Published 2025-12-10.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81342

Official notice

2025-12-10

The hazard

The Vanta A71200 CP app V2.0.2465 and V2.0.2683 will display a "system error" if the user selects the start usage button twice when interrogating a new implantable neurostimulator (INS). This error prompts the user to restart the CP app and contact Medtronic for assistance. Upon subsequent interrogation of the INS, the Vanta A71200 CP app will display a "validation error" prompting the user to resolve the error in order to proceed with the INS interrogation. If the validation error is not resolved, the user is unable to complete the first attempt at interrogating the INS.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

A71200

Other

RA RA-81342

Model Numbers

A71200

Recall Numbers

81342

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Vanta/Sequentia Lt Clinician Programmer Application, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - A71200

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - A71200

Risk / hazard

The Vanta A71200 CP app V2.0.2465 and V2.0.2683 will display a "system error" if the user selects the start usage button twice when interrogating a new implantable neurostimulator (INS). This error prompts the user to restart the CP app and contact Medtronic for assistance. Upon subsequent interrogation of the INS, the Vanta A71200 CP app will display a "validation error" prompting the user to resolve the error in order to proceed with the INS interrogation. If the validation error is not resolved, the user is unable to complete the first attempt at interrogating the INS.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Medtronic

Manufacturer

Not provided

Phone

Not provided

Brand: Medtronic.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

81342

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Synchromed clinician programmer application

Critical riskHealth Canada15 Apr 2026
CA
Recall image placeholder

Octopus nuvo, octopus evo as and octopus 4

High riskHealth Canada13 May 2026
CA
Recall image placeholder

Harmony™ delivery catheter system

Medium riskHealth Canada4 Jun 2026
CA
Recall image placeholder

Medtronic dlp retrograde cardioplegia cannulae

Medium riskHealth Canada30 Apr 2026
CA
Recall image placeholder

Dlp select 3d arterial cannula and dlp elongated one piece arterial cannulae

Medium riskHealth Canada3 Jul 2026
CA
Recall image placeholder

Medtronic restore™ clinician programmer application

Medium riskHealth Canada28 Nov 2025
CA
Recall image placeholder

Carelink smartsync™ applications

Critical riskHealth Canada25 Jun 2025
CA
Recall image placeholder

Inceptiv

Critical riskHealth Canada10 Jun 2025
CA
Recall image placeholder

Dlp cardiac vents & sumps

Medium riskHealth Canada24 Sept 2025
CA
Recall image placeholder

Aurora ev-icd™ mri surescan™

Medium riskHealth Canada6 Nov 2025
CA
Recall image placeholder

Becker and exacta external drainage and monitoring systems

Medium riskHealth Canada14 Nov 2024
CA
Recall image placeholder

Carelink programmer

Medium riskHealth Canada26 Aug 2024

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 10 Dec 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.