No image published for this recall

No image published for this recall

Liko universal twinbar 670 qrh

Official notice

7 Mar 2022

Last checked

7 Mar 2022

Medium riskHealth Canada

An official Health Canada notice concerns Liko universal twinbar 670 qrh. During manufacturing of universal twinbar 670 Qrh, the wrong center bolt was used for some twinbars. The wrong bolt does, under some circumstances, prevent a rotation of the twinbar around its center bolt. When the rotation around the bolt is inhibited, a momentum is transferred to the attachment between the twinbar and the lift. This has caused breakage in the field resulting in patient falls. To date, there have been no reported injuries. Recall start date: Feb 25. 2022 Published 2022-03-07.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63968

Official notice

2022-03-07

The hazard

During manufacturing of universal twinbar 670 Qrh, the wrong center bolt was used for some twinbars. The wrong bolt does, under some circumstances, prevent a rotation of the twinbar around its center bolt. When the rotation around the bolt is inhibited, a momentum is transferred to the attachment between the twinbar and the lift. This has caused breakage in the field resulting in patient falls. To date, there have been no reported injuries. Recall start date: Feb 25. 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

63968

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected Products - Liko Universal Twinbar 670 Qrh, Lot or Serial Number - Not applicable., Model or catalog number - 3156087, P3156087

Risk / hazard

During manufacturing of universal twinbar 670 Qrh, the wrong center bolt was used for some twinbars. The wrong bolt does, under some circumstances, prevent a rotation of the twinbar around its center bolt. When the rotation around the bolt is inhibited, a momentum is transferred to the attachment between the twinbar and the lift. This has caused breakage in the field resulting in patient falls. To date, there have been no reported injuries. Recall start date: Feb 25. 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Liko, Liko Universal Twinbar 670 Qrh.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

63968

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 7 Mar 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.