Lifepak ® 35 monitor defibrillator

Official notice

4 Sept 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Lifepak ® 35 monitor defibrillator. During production of LP35, Stryker identified that certain printer and modem access port connectors may not have been fully seated within the rear housing assembly. As a result, the connectors may appear flush with or recessed relative to the retaining nut rather than extending beyond it as intended. Published 2026-09-04.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82601

Official notice

2026-09-04

The hazard

During production of LP35, Stryker identified that certain printer and modem access port connectors may not have been fully seated within the rear housing assembly. As a result, the connectors may appear flush with or recessed relative to the retaining nut rather than extending beyond it as intended.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

70335-000180, 99335-000123, 99335-000124, 70335-000182, 70335-000185, 70335-000184, 99335-000121, 99335-000126, 70335-000183, 99335-000035, 99335-000125, 70335-000047, 70335-000181, 99335-000122

Other

RA RA-82601

Model numbers

70335-000180, 99335-000123, 99335-000124, 70335-000182, 70335-000185, 70335-000184, 99335-000121, 99335-000126, 70335-000183, 99335-000035, 99335-000125, 70335-000047, 70335-000181, 99335-000122

Recall Numbers

82601

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000180

Product 2

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000123

Product 3

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000124

Product 4

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000182

Product 5

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000185

Product 6

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000184

Product 7

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000121

Product 8

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000126

Product 9

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000183

Product 10

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000035

Product 11

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000125

Product 12

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000047

Product 13

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 70335-000181

Product 14

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 99335-000122

Hazard

During production of LP35, Stryker identified that certain printer and modem access port connectors may not have been fully seated within the rear housing assembly. As a result, the connectors may appear flush with or recessed relative to the retaining nut rather than extending beyond it as intended.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Physio-control.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82601

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Sept 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button