Hamilton medical ag breathing set mri

Official notice

28 May 2026

Last checked

28 May 2026

Medium riskHealth Canada
Watch Hamilton Medical

An official Health Canada notice concerns Hamilton medical ag breathing set mri. Certain lots of the expiratory valve set pre-assembled to the breathing circuit set, coaxial (see FSN) do not perform as intended. In some cases, the expiratory valve membrane adhered to the expiratory valve body's sealing ring, thereby impairing valve opening and expiratory gas flow. The issue is not detected during the pre-operative test of the ventilator unless a test lung is actively ventilated. In case of an adhering expiratory valve membrane, the malfunction appears within the first breaths after ventilation is initiated. When the expiratory gas flow is impeded, the end-expiratory pressure is too high, or the end-expiratory flow is too low. Consequently, the ventilator triggers the high priority alarm "Exhalation Obstructed". The obstructed air is released through the device, e.g.the obstruction valve opens so that the patient can exhale through the inspiratory limb. Published 2026-05-28.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82133

Official notice

2026-05-28

The hazard

Certain lots of the expiratory valve set pre-assembled to the breathing circuit set, coaxial (see FSN) do not perform as intended. In some cases, the expiratory valve membrane adhered to the expiratory valve body's sealing ring, thereby impairing valve opening and expiratory gas flow. The issue is not detected during the pre-operative test of the ventilator unless a test lung is actively ventilated. In case of an adhering expiratory valve membrane, the malfunction appears within the first breaths after ventilation is initiated. When the expiratory gas flow is impeded, the end-expiratory pressure is too high, or the end-expiratory flow is too low. Consequently, the ventilator triggers the high priority alarm "Exhalation Obstructed". The obstructed air is released through the device, e.g.the obstruction valve opens so that the patient can exhale through the inspiratory limb.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

260167, 260168

Other

RA RA-82133

Model Numbers

260167, 260168

Recall Numbers

82133

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 260167

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 260168

Risk / hazard

Certain lots of the expiratory valve set pre-assembled to the breathing circuit set, coaxial (see FSN) do not perform as intended. In some cases, the expiratory valve membrane adhered to the expiratory valve body's sealing ring, thereby impairing valve opening and expiratory gas flow. The issue is not detected during the pre-operative test of the ventilator unless a test lung is actively ventilated. In case of an adhering expiratory valve membrane, the malfunction appears within the first breaths after ventilation is initiated. When the expiratory gas flow is impeded, the end-expiratory pressure is too high, or the end-expiratory flow is too low. Consequently, the ventilator triggers the high priority alarm "Exhalation Obstructed". The obstructed air is released through the device, e.g.the obstruction valve opens so that the patient can exhale through the inspiratory limb.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Hamilton Medical

Manufacturer

Not provided

Phone

Not provided

Brand: Hamilton Medical.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

82133

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Intellicuff®

Medium riskHealth Canada3 Jun 2026
CA
Recall image placeholder

Hamilton-c2 intensive care ventilator

High riskHealth Canada19 Jan 2026
CA
Recall image placeholder

Hamilton-c6 intensive care ventilator

Critical riskHealth Canada11 Jun 2025
CA
Recall image placeholder

In2flow nasal cannula, with od15 adapter

Critical riskHealth Canada26 Feb 2025
CA
Recall image placeholder

Hamilton c2 and c3 intensive care ventilators

High riskHealth Canada17 Feb 2025
CA
Recall image placeholder

In2flow nasal cannulas

Critical riskHealth Canada28 Sept 2023
CA
Recall image placeholder

Hamilton-c2 c3 c1 t1 ventilators

High riskHealth Canada26 Sept 2023
CA
Recall image placeholder

Hamilton-c1 and hamilton-t1 ventilators

High riskHealth Canada21 Aug 2023
CA
Recall image placeholder

Hamilton c6 intensive care ventilator

Medium riskHealth Canada23 Mar 2023
CA
Recall image placeholder

Hamilton-c6 intensive care ventilator

High riskHealth Canada29 May 2024
CA
Recall image placeholder

Hamilton c6 intensive care ventilator

Medium riskHealth Canada7 Jun 2022
CA
Recall image placeholder

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Medium riskHealth Canada23 Jul 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 28 May 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.