Hamilton-c6 intensive care ventilator

Official notice

11 Jun 2025

Last checked

11 Jun 2025

Critical riskHealth Canada
Watch Hamilton Medical

An official Health Canada notice concerns Hamilton-c6 intensive care ventilator. Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state. Recall start date: June 2, 2025 Published 2025-06-11.

What to do now

The local approved distribution partner is the first point of contact.

Recall number

77583

Official notice

2025-06-11

The hazard

Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state. Recall start date: June 2, 2025

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

77583

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Hamilton-C6 Intensive Care Ventilator, Lot or serial number - Serial numbers from 13636 till 13950 Serial numbers from 13966 till 14753, Model or catalogue number - 160021

Risk / hazard

Hamilton Medical AG received information from the supplier of the ESM-board which is installed in the Hamilton-C6, indicating that certain ESM-board may contain a mechanically damaged ceramic capacitor. Since only certain serial numbers (from 13636 till 13950 and from 13966 till 14753) of the Hamilton-C6 are affected with the defective ESM-board, this FSN is limited to these devices defined above. Damage to the capacitor could, over time, lead to a loss of its function due to oxidation or similar degradation processes. The worst-case scenario is a local short circuit of the aforementioned capacitor, with different outcomes depending on the device state. Recall start date: June 2, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Hamilton Medical, Hamilton-c6 Intensive Care Ventilator

Manufacturer

Not provided

Phone

Not provided

Brand: Hamilton Medical, Hamilton-c6 Intensive Care Ventilator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

77583

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Hamilton medical ag breathing set mri

Medium riskHealth Canada28 May 2026
CA
Recall image placeholder

Intellicuff®

Medium riskHealth Canada3 Jun 2026
CA
Recall image placeholder

Hamilton-c2 intensive care ventilator

High riskHealth Canada19 Jan 2026
CA
Recall image placeholder

In2flow nasal cannula, with od15 adapter

Critical riskHealth Canada26 Feb 2025
CA
Recall image placeholder

Hamilton c2 and c3 intensive care ventilators

High riskHealth Canada17 Feb 2025
CA
Recall image placeholder

Hamilton-c6 intensive care ventilator

High riskHealth Canada29 May 2024
CA
Recall image placeholder

In2flow nasal cannulas

Critical riskHealth Canada28 Sept 2023
CA
Recall image placeholder

Hamilton-c2 c3 c1 t1 ventilators

High riskHealth Canada26 Sept 2023
CA
Recall image placeholder

Hamilton-c1 and hamilton-t1 ventilators

High riskHealth Canada21 Aug 2023
CA
Recall image placeholder

Hamilton c6 intensive care ventilator

Medium riskHealth Canada23 Mar 2023
CA
Recall image placeholder

Hamilton c6 intensive care ventilator

Medium riskHealth Canada7 Jun 2022
CA
Recall image placeholder

Intramedullary bone saw blade assembly

Medium riskHealth Canada3 Jul 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 Jun 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.