Hamilton-c2 c3 c1 t1 ventilators
Official notice
26 Sept 2023
Last checked
26 Sept 2023
An official Health Canada notice concerns Hamilton-c2 c3 c1 t1 ventilators. We have become aware of a malfunction with the Hamilton-C2/C3/C1/T1 ventilators when used long term in neonatal patient group. If an affected device is used accumulatively for 91 days in neonatal patient group without a restart of the device, the device will switch to "ambient state" and therefore stop ventilation. During the "ambient state" the device is alarming audibly and visually until it is switched off by the user. Hence, it is mandatory to apply a periodic preventive restart of affected devices. Devices and software version affected: All hamilton-C1/T1 with software version ≤2.2.10 and neonatal option installed All hamilton-C2 with software version ≤2.2.5 and neonatal option installed All hamilton-C3 with software version ≤2.0.9 and neonatal option installed Recall start date: June 1, 2023 Published 2023-09-26.
What to do now
Contact the manufacturer if you require additional information.
Recall number
73858
Official notice
2023-09-26
The hazard
The hazard
We have become aware of a malfunction with the Hamilton-C2/C3/C1/T1 ventilators when used long term in neonatal patient group. If an affected device is used accumulatively for 91 days in neonatal patient group without a restart of the device, the device will switch to "ambient state" and therefore stop ventilation. During the "ambient state" the device is alarming audibly and visually until it is switched off by the user. Hence, it is mandatory to apply a periodic preventive restart of affected devices. Devices and software version affected: All hamilton-C1/T1 with software version ≤2.2.10 and neonatal option installed All hamilton-C2 with software version ≤2.2.5 and neonatal option installed All hamilton-C3 with software version ≤2.0.9 and neonatal option installed Recall start date: June 1, 2023
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Hamilton Medical, Hamilton-c2 C3 C1 T1 Ventilators.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Intellicuff®

Hamilton medical ag breathing set mri

Hamilton-c2 intensive care ventilator

Hamilton-c6 intensive care ventilator

In2flow nasal cannula, with od15 adapter

Hamilton c2 and c3 intensive care ventilators

Hamilton-c6 intensive care ventilator

In2flow nasal cannulas

Hamilton-c1 and hamilton-t1 ventilators

Hamilton c6 intensive care ventilator

Hamilton c6 intensive care ventilator

Aisys, avance, and aespire anesthesia machines
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 26 Sept 2023. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.