Open official notice

Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and or limbs

Official notice

7 Aug 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and or limbs. As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne Published 2013-08-07.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

34,617

Recall number

Z-1836-2013

Official notice

2013-08-07

The hazard

As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63108

Recall Numbers

Z-1836-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.

Units affected

34,617

Hazard

As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CO · Colombia

Affected area 10

CY · Cyprus

Affected area 11

CZ · Czech Republic

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

ES · Spain

Affected area 15

FI · Finland

Affected area 16

FR · France

Affected area 17

GB · United Kingdom

Affected area 18

GR · Greece

Affected area 19

IE · Ireland

Affected area 20

IL · Israel

Affected area 21

IN · India

Affected area 22

IT · Italy

Affected area 23

JP · Japan

Affected area 24

MX · Mexico

Affected area 25

NL · Netherlands

Affected area 26

SE · Sweden

Affected area 27

TR · Turkey

Affected area 28

ZA · South Africa

Company & contacts

Manufacturer

St. Jude Medical

Phone

Not provided

Brand: St Jude Medical. Contact: St. Jude Medical.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1836-2013

Event ID

63108

Units affected

34,617

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 7 Aug 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button