An official FDA notice concerns Cortical Screw for M DN Intramedullary Fixation. Metaphyseal Diaphyseal Nail System, Orthopedic surgical device. The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing. Published 2015-06-10.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

289,616

Recall number

Z-1694-2015

Official notice

2015-06-10

The hazard

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

71243

Recall Numbers

Z-1694-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Cortical Screw for M/DN Intramedullary Fixation. Metaphyseal/Diaphyseal Nail System, Orthopedic surgical device.

Units affected

289,616

Hazard

The entire scope of 4.5mm Cortical Screws listed as compatible with the M/DN system is not compatible with the M/DN system. This has the potential to lead to a situation in which the screw would bind upon entering the nail and have to be removed during the surgery or removal of the screw after fracture healing.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

BE · Belgium

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CH · Switzerland

Affected area 7

CN · China

Affected area 8

DE · Germany

Affected area 9

ES · Spain

Affected area 10

FR · France

Affected area 11

GB · United Kingdom

Affected area 12

IN · India

Affected area 13

IT · Italy

Affected area 14

JO · Jordan

Affected area 15

JP · Japan

Affected area 16

KR · South Korea

Affected area 17

LT · Lithuania

Affected area 18

MT · Malta

Affected area 19

MX · Mexico

Affected area 20

MY · Malaysia

Affected area 21

RS · Serbia

Affected area 22

SA · Saudi Arabia

Affected area 23

SG · Singapore

Affected area 24

TH · Thailand

Affected area 25

TW · Taiwan

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1694-2015

Event ID

71243

Units affected

289,616

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Jun 2015. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button