Open official notice

Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius

Official notice

17 Dec 2014

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius. Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes. Published 2014-12-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

14

Recall number

Z-0574-2015

Official notice

2014-12-17

The hazard

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

L75066

Event Ids

69613

Recall Numbers

Z-0574-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial Spinal System, used for posterior lumbar fusion procedures.

Units affected

14

Hazard

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0574-2015

Event ID

69613

Units affected

14

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Dec 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button