An official FDA notice concerns Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452. Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay. Published 2019-08-21.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Recall number

Z-2282-2019

Official notice

2019-08-21

The hazard

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

83470

Recall Numbers

Z-2282-2019

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Aries Bordetella Assay, REF 50-10037, UDI # 00840487101452

Hazard

Through internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AK

Location 2

US · United States · AL

Location 3

US · United States · AZ

Location 4

US · United States · CA

Location 5

US · United States · CO

Location 6

US · United States · FL

Location 7

US · United States · GA

Location 8

US · United States · IL

Location 9

US · United States · KS

Location 10

US · United States · KY

Location 11

US · United States · LA

Location 12

US · United States · MA

Location 13

US · United States · MD

Location 14

US · United States · ME

Location 15

US · United States · MI

Location 16

US · United States · MN

Location 17

US · United States · MO

Location 18

US · United States · MS

Location 19

US · United States · MT

Location 20

US · United States · NC

Location 21

US · United States · NE

Location 22

US · United States · NJ

Location 23

US · United States · NM

Location 24

US · United States · NY

Location 25

US · United States · OH

Location 26

US · United States · OR

Location 27

US · United States · PA

Location 28

US · United States · TN

Location 29

US · United States · TX

Location 30

US · United States · UT

Location 31

US · United States · VA

Location 32

US · United States · VT

Location 33

US · United States · WA

Location 34

US · United States · WI

Affected area 1

US · United States

Affected area 2

BE · Belgium

Affected area 3

CA · Canada

Affected area 4

DE · Germany

Affected area 5

GB · United Kingdom

Affected area 6

SG · Singapore

Affected area 7

TH · Thailand

Company & contacts

Brand

Manufacturer

Luminex Corporation

Phone

Not provided

Brand: Luminex. Contact: Luminex Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2282-2019

Event ID

83470

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Aug 2019. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button