Open official notice

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

Official notice

11 Nov 2026

Last checked

17 Nov 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303. Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection. Published 2026-11-11.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

30,168

Recall number

Z-0359-2027

Official notice

2026-11-11

The hazard

Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

100036

Recall Numbers

Z-0359-2027

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

LIAISON PLEX Gram-Negative Blood Culture Assay Catalog No. 20-005-303

Units affected

30,168

Hazard

Instructions for Use updated to include explicit guidance that repeat testing is recommended whenever a positive family-level only result is obtained. Issue may contribute to incorrect or delayed diagnosis of bloodstream infection.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Luminex Corporation

Phone

Not provided

Brand: Luminex. Contact: Luminex Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0359-2027

Event ID

100036

Units affected

30,168

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 11 Nov 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button