Image not available in RecallScope.
Tunneloc round head interference screw 8mmx25mm biomet sports medicine manufactures a variety of internal fixation devices intended to aid
Official notice
6 Feb 2013
An official FDA notice concerns Tunneloc round head interference screw 8mmx25mm biomet sports medicine manufactures a variety of internal fixation devices intended to aid. Thru hole at the tip of the screw is undersized. Published 2013-02-06.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
44
Recall number
Z-0753-2013
Official notice
2013-02-06
The hazard
The hazard
Thru hole at the tip of the screw is undersized.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Location 4
Location 5
Location 6
Location 7
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Company & contacts
Company & contacts
Brand: Biomet. Contact: Biomet, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.