PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument

Official notice

10 Jun 2015

RecallScope risk estimate:MediumFDA

An official FDA notice concerns PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument. Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response. Published 2015-06-10.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,689

Recall number

Z-1696-2015

Official notice

2015-06-10

The hazard

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

71250

Recall Numbers

Z-1696-2015

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

PRI Femoral Impactor Block for knee prosthesis surgery. Orthopedic Manual Surgical Instrument

Units affected

1,689

Hazard

Potential for uncured adhesive between the metal threaded insert and the impactor block to leak out of the assembled part following the autoclave sterilization process. The Master Bond epoxy serves as a secondary locking mechanism to fix the metal insert to the impactor head. Foreign material has been observed on the impactor block. This has the potential for a biological response.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CZ · Czech Republic

Affected area 9

DE · Germany

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

IL · Israel

Affected area 14

IN · India

Affected area 15

IS · Iceland

Affected area 16

IT · Italy

Affected area 17

JP · Japan

Affected area 18

KR · South Korea

Affected area 19

MY · Malaysia

Affected area 20

NO · Norway

Affected area 21

SA · Saudi Arabia

Affected area 22

SE · Sweden

Affected area 23

SG · Singapore

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1696-2015

Event ID

71250

Units affected

1,689

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Jun 2015. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button