NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee

Official notice

5 Dec 2012

RecallScope risk estimate:MediumFDA

An official FDA notice concerns NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee. Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset Published 2012-12-05.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

123

Recall number

Z-0437-2013

Official notice

2012-12-05

The hazard

Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63544

Recall Numbers

Z-0437-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen¿¿ system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.

Units affected

123

Hazard

Zimmer is initiating a lot specific recall of the NexGen¿ Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AZ

Location 2

US · United States · CA

Location 3

US · United States · FL

Location 4

US · United States · IA

Location 5

US · United States · IL

Location 6

US · United States · IN

Location 7

US · United States · MO

Location 8

US · United States · NE

Location 9

US · United States · NJ

Location 10

US · United States · NY

Location 11

US · United States · OH

Location 12

US · United States · PA

Location 13

US · United States · WA

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

DE · Germany

Affected area 4

FR · France

Affected area 5

GB · United Kingdom

Affected area 6

HU · Hungary

Affected area 7

NL · Netherlands

Affected area 8

SE · Sweden

Affected area 9

TH · Thailand

Company & contacts

Brand

Manufacturer

Zimmer, Inc.

Phone

Not provided

Brand: Zimmer. Contact: Zimmer, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0437-2013

Event ID

63544

Units affected

123

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Dec 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button