Open official notice

Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack 6416-100 6416-1005

Official notice

20 Jul 2016

RecallScope risk estimate:CriticalFDA

Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack 6416-100 6416-1005 6416-140 6416-1405 6416-200 6416-2005 For temporary intracardiac pacing and EGM recording. The system is disposable, for temporary single patient use, with a contemplated implant duration of 7 days or less. The system is used with Medtronic temporary external pacemakers.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

11,697

Recall number

Z-2200-2016

Official notice

2016-07-20

The hazard

This Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

74424

Recall Numbers

Z-2200-2016

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack 6416-100 6416-1005 6416-140 6416-1405 6416-200 6416-2005 For temporary intracardiac pacing and EGM recording. The system is disposable, for temporary single patient use, with a contemplated implant duration of 7 days or less. The system is used with Medtronic temporary external pacemakers.

Units affected

11,697

Hazard

This Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AU · Australia

Affected area 4

BG · Bulgaria

Affected area 5

CA · Canada

Affected area 6

CH · Switzerland

Affected area 7

CO · Colombia

Affected area 8

DE · Germany

Affected area 9

ES · Spain

Affected area 10

FR · France

Affected area 11

GB · United Kingdom

Affected area 12

HK · Hong Kong

Affected area 13

HU · Hungary

Affected area 14

IE · Ireland

Affected area 15

IL · Israel

Affected area 16

IT · Italy

Affected area 17

JO · Jordan

Affected area 18

NL · Netherlands

Affected area 19

NO · Norway

Affected area 20

PL · Poland

Affected area 21

PT · Portugal

Affected area 22

RS · Serbia

Affected area 23

SA · Saudi Arabia

Affected area 24

SE · Sweden

Affected area 25

SI · Slovenia

Affected area 26

TR · Turkey

Affected area 27

ZA · South Africa

Company & contacts

Manufacturer

Medtronic Inc., Cardiac Rhythm and Heart Failure

Phone

Not provided

Brand: Medtronic Cardiac Rhythm And Heart Failure. Contact: Medtronic Inc., Cardiac Rhythm and Heart Failure.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2200-2016

Event ID

74424

Units affected

11,697

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 20 Jul 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button