Open official notice

Bd max dna mmk (spc) for laboratory use, catalog no. 442829

Official notice

9 May 2018

RecallScope risk estimate:LowFDA

An official FDA notice concerns Bd max dna mmk (spc) for laboratory use, catalog no. 442829. The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay. Published 2018-05-09.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

13,898

Recall number

Z-1544-2018

Official notice

2018-05-09

The hazard

The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

79769

Recall Numbers

Z-1544-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

BD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829

Units affected

13,898

Hazard

The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · FL

Location 3

US · United States · NJ

Location 4

US · United States · NY

Location 5

US · United States · OR

Location 6

US · United States · PA

Location 7

US · United States · VA

Location 8

US · United States · WI

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AU · Australia

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CL · Chile

Affected area 7

CN · China

Affected area 8

CO · Colombia

Affected area 9

HK · Hong Kong

Affected area 10

JP · Japan

Affected area 11

SG · Singapore

Affected area 12

TH · Thailand

Affected area 13

TW · Taiwan

Company & contacts

Manufacturer

Becton Dickinson & Co.

Phone

Not provided

Brand: Becton Dickinson. Contact: Becton Dickinson & Co..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class III

Recall details

Recall number

Z-1544-2018

Event ID

79769

Units affected

13,898

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 May 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button