Open official notice

The bd max mrsa assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of

Official notice

17 Sept 2014

RecallScope risk estimate:MediumFDA

An official FDA notice concerns The bd max mrsa assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of. The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results. Published 2014-09-17.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

585,048

Recall number

Z-2618-2014

Official notice

2014-09-17

The hazard

The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

68991

Recall Numbers

Z-2618-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.

Units affected

585,048

Hazard

The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CA

Location 2

US · United States · FL

Location 3

US · United States · HI

Location 4

US · United States · IL

Location 5

US · United States · IN

Location 6

US · United States · MD

Location 7

US · United States · MI

Location 8

US · United States · MN

Location 9

US · United States · MT

Location 10

US · United States · NC

Location 11

US · United States · ND

Location 12

US · United States · NJ

Location 13

US · United States · NM

Location 14

US · United States · NV

Location 15

US · United States · NY

Location 16

US · United States · OH

Location 17

US · United States · OK

Location 18

US · United States · OR

Location 19

US · United States · PA

Location 20

US · United States · SC

Location 21

US · United States · TN

Location 22

US · United States · TX

Location 23

US · United States · UT

Location 24

US · United States · VA

Location 25

US · United States · WA

Location 26

US · United States · WI

Affected area 1

US · United States

Company & contacts

Manufacturer

Becton Dickinson & Co.

Phone

Not provided

Brand: Becton Dickinson. Contact: Becton Dickinson & Co..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2618-2014

Event ID

68991

Units affected

585,048

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Sept 2014. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button