Open official notice

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening

Official notice

17 Jul 2024

RecallScope risk estimate:CriticalFDA

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

14,885

Recall number

Z-2324-2024

Official notice

2024-07-17

The hazard

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

94792

Recall Numbers

Z-2324-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardiac Arrest (SCA). Model/Catalog Number: G5S-00A G5S-00A-TSO G5S-00C G5S-01A G5S-02A G5S-02C G5S-02-L G5S-04A G5S-05A G5S-05C G5S-06A G5S-06C G5S-08A G5S-08C G5S-10A G5S-10C G5S-11A G5S-11C G5S-12C G5S-13A G5S-14A G5S-15A G5S-17A G5S-17C G5S-19A G5S-23C G5S-29A G5S-29C G5S-31A G5S-31A-SJA G5S-31C-SJA G5S-36A G5S-37A G5S-41A G5S-41C G5S-80A G5S-80A-TSO G5S-80C G5S-80-L G5S-82A G5S-82C G5S-83C G5S-90C

Units affected

14,885

Hazard

G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Zoll Medical Corporation

Phone

Not provided

Brand: Zoll Medical. Contact: Zoll Medical Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2324-2024

Event ID

94792

Units affected

14,885

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jul 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.