Open official notice

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35

Official notice

19 Apr 2017

RecallScope risk estimate:MediumFDA

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

6,324

Recall number

Z-1765-2017

Official notice

2017-04-19

The hazard

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

1570, 1580, 1590

Event Ids

76665

Recall Numbers

Z-1765-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITROS Immunodiagnostic Products NT-pro BNP Reagent Pack, REF/Catalog 680 2156, IVD --- Ortho-Clinical Diagnostics Pencoed, Bridgend CF35 5PZ, UK

Units affected

6,324

Hazard

Increased frequency of calibration failures for VITROS Immunodiagnostic Products NT-proBNP Reagent Lots 1568, 1570, 1580 and 1590, due to background signals for these affected NT-proBNP lots that have been increasing with time since release testing, resulting in higher than expected VITROS Immunodiagnostic Products NT-proBNP level 1 calibrator signals produced, leading to calibration failures.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

BR · Brazil

Affected area 4

CA · Canada

Affected area 5

CL · Chile

Affected area 6

CN · China

Affected area 7

CO · Colombia

Affected area 8

DE · Germany

Affected area 9

ES · Spain

Affected area 10

FR · France

Affected area 11

GB · United Kingdom

Affected area 12

IN · India

Affected area 13

IT · Italy

Affected area 14

JP · Japan

Affected area 15

MX · Mexico

Affected area 16

PL · Poland

Affected area 17

SG · Singapore

Affected area 18

VE · Venezuela

Company & contacts

Manufacturer

ORTHO-CLINICAL DIAGNOSTICS

Phone

Not provided

Brand: Ortho-clinical Diagnostics. Contact: ORTHO-CLINICAL DIAGNOSTICS.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1765-2017

Event ID

76665

Units affected

6,324

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Apr 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button