Open official notice

VITROS 3600 Immunodiagnostic System (New and Refurbished)

Official notice

13 Mar 2024

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VITROS 3600 Immunodiagnostic System (New and Refurbished). Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erroneous test results. Published 2024-03-13.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

489

Recall number

Z-1220-2024

Official notice

2024-03-13

The hazard

Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

93892

Recall Numbers

Z-1220-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITROS 3600 Immunodiagnostic System (New and Refurbished)

Units affected

489

Hazard

Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS¿ 3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AU · Australia

Affected area 3

BE · Belgium

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CL · Chile

Affected area 7

CN · China

Affected area 8

CO · Colombia

Affected area 9

DE · Germany

Affected area 10

DK · Denmark

Affected area 11

ES · Spain

Affected area 12

FR · France

Affected area 13

GB · United Kingdom

Affected area 14

IN · India

Affected area 15

IT · Italy

Affected area 16

JP · Japan

Affected area 17

MX · Mexico

Affected area 18

NL · Netherlands

Affected area 19

NO · Norway

Affected area 20

PT · Portugal

Affected area 21

RU · Russia

Affected area 22

SE · Sweden

Affected area 23

SG · Singapore

Company & contacts

Manufacturer

Ortho-Clinical Diagnostics, Inc.

Phone

Not provided

Brand: Ortho-clinical Diagnostics. Contact: Ortho-Clinical Diagnostics, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1220-2024

Event ID

93892

Units affected

489

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Mar 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button