Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b

Official notice

25 Jul 2018

RecallScope risk estimate:CriticalFDA

Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase¿ Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase¿ Cooled Laser Applicator System 15mm Tip Product Usage: The Visualase Thermal Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for wavelengths 800 nm - 1060 nm. When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the Visualase system can process images to determine relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,283

Recall number

Z-2501-2018

Official notice

2018-07-25

The hazard

Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

80481

Recall Numbers

Z-2501-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase¿ Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase¿ Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase¿ Cooled Laser Applicator System 15mm Tip Product Usage: The Visualase Thermal Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for wavelengths 800 nm - 1060 nm. When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the Visualase system can process images to determine relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.

Units affected

1,283

Hazard

Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

IL · Israel

Company & contacts

Manufacturer

Medtronic Navigation, Inc.

Phone

Not provided

Brand: Medtronic Navigation. Contact: Medtronic Navigation, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2501-2018

Event ID

80481

Units affected

1,283

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 25 Jul 2018. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button