Image not available in RecallScope.
Vista Dental Products Chlor-XTRA - 12mL Prefilled Syringes, QTY: 8, REF 503850
Official notice
9 Oct 2019
An official FDA notice concerns Vista Dental Products Chlor-XTRA - 12mL Prefilled Syringes, QTY: 8, REF 503850. The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution. Published 2019-10-09.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
384
Recall number
Z-0017-2020
Official notice
2019-10-09
The hazard
The hazard
The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Company & contacts
Company & contacts
Brand: Inter-med. Contact: Inter-Med Llc.
Source & status
Source & status
Recall details
Recall details
Recall number
Event id
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest