Open official notice

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Official notice

9 Aug 2023

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF). Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema. Published 2023-08-09.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

148,941

Recall number

Z-2247-2023

Official notice

2023-08-09

The hazard

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 05050474700901

Event Ids

92539

Recall Numbers

Z-2247-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Units affected

148,941

Hazard

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BE · Belgium

Affected area 7

BG · Bulgaria

Affected area 8

BR · Brazil

Affected area 9

CA · Canada

Affected area 10

CH · Switzerland

Affected area 11

CL · Chile

Affected area 12

CN · China

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

DK · Denmark

Affected area 16

EE · Estonia

Affected area 17

EG · Egypt

Affected area 18

ES · Spain

Affected area 19

FI · Finland

Affected area 20

FR · France

Affected area 21

GB · United Kingdom

Affected area 22

GR · Greece

Affected area 23

HK · Hong Kong

Affected area 24

IL · Israel

Affected area 25

IN · India

Affected area 26

IT · Italy

Affected area 27

JO · Jordan

Affected area 28

JP · Japan

Affected area 29

KR · South Korea

Affected area 30

KW · Kuwait

Affected area 31

MY · Malaysia

Affected area 32

NL · Netherlands

Affected area 33

NO · Norway

Affected area 34

NZ · New Zealand

Affected area 35

PE · Peru

Affected area 36

PL · Poland

Affected area 37

PT · Portugal

Affected area 38

SA · Saudi Arabia

Affected area 39

SE · Sweden

Affected area 40

SG · Singapore

Affected area 41

TH · Thailand

Affected area 42

TR · Turkey

Affected area 43

TW · Taiwan

Affected area 44

VE · Venezuela

Affected area 45

ZA · South Africa

Company & contacts

Manufacturer

Johnson & Johnson Surgical Vision, Inc.

Phone

Not provided

Brand: Johnson Johnson Surgical Vision. Contact: Johnson & Johnson Surgical Vision, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2247-2023

Event ID

92539

Units affected

148,941

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Aug 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button