An official FDA notice concerns Verify¿ Bowie-Dick Test Card, Model #S3098, Steris Biological Operations, 9325 Pinecone Drive, Mentor, OH 44060 The Verify Bowie-Dick Test. STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges. Published 2012-09-26.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

629

Recall number

Z-2417-2012

Official notice

2012-09-26

The hazard

STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

63129

Recall Numbers

Z-2417-2012

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Verify¿ Bowie-Dick Test Card, Model #S3098, Steris Biological Operations, 9325 Pinecone Drive, Mentor, OH 44060 The Verify Bowie-Dick Test Card consists of a card printed on one side with bars of chemical indicator ink. On the opposite side of the card, space is provided for critical cycle information to be recorded. The card is laminated inside a thin clear plastic enclosure. The enclosure has two holes in it to allow air removal and steam penetration. The preassembled test is used to evaluate the effectiveness of air removal from the sterilizer chamber during prevacuum pulse steam sterilizer cycle. Following a prevacuum cycle, the chemical bars of the Bowie-Dick Test card uniformly darken indicating that residual air has been sufficiently removed to allow complete steam penetration into the test card. If air is trapped in the card during the exposure phase of the cycle, the color change of the bars will be incomplete or uneven. Thus the card can provide an immediate indication of inadequate removal of air during a cycle. The Bowie-Dick Type test is performed in an empty chamber each day the sterilizer is to be used, usually before the first sterilization cycle.

Units affected

629

Hazard

STERIS has learned that the Verify Bowie Dick Test Cards are not performing to product specifications. Specifically, test cards have resulted in false fail results when sterilizer performance is within acceptable ranges.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Brand

Manufacturer

Steris Corporation

Phone

Not provided

Brand: Steris. Contact: Steris Corporation.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2417-2012

Event ID

63129

Units affected

629

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 26 Sept 2012. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button