VariSoft infusion set, Model Number 1002827

Official notice

29 Nov 2023

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns VariSoft infusion set, Model Number 1002827. The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. Which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis. Published 2023-11-29.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

8,028

Recall number

Z-0232-2024

Official notice

2023-11-29

The hazard

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. Which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

93269

Recall Numbers

Z-0232-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VariSoft infusion set, Model Number 1002827

Units affected

8,028

Hazard

The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. Which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Unomedical A/S

Phone

Not provided

Brand: Unomedical A S. Contact: Unomedical A/S.

Source & status

Source

Tier

official

Category

ELECTRONICS

Official classification

Class I

Recall details

Recall number

Z-0232-2024

Event id

93269

Units affected

8,028

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 29 Nov 2023. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button