Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and

Official notice

9 Feb 2022

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and. A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy. Published 2022-02-09.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

10

Recall number

Z-0532-2022

Official notice

2022-02-09

The hazard

A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

89335

Recall Numbers

Z-0532-2022

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter)

Units affected

10

Hazard

A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · AR

Location 2

US · United States · IN

Location 3

US · United States · KS

Location 4

US · United States · MA

Location 5

US · United States · MO

Location 6

US · United States · NE

Location 7

US · United States · WV

Affected area 1

US · United States

Company & contacts

Manufacturer

LivaNova USA, Inc.

Phone

Not provided

Brand: Livanova. Contact: LivaNova USA, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0532-2022

Event ID

89335

Units affected

10

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Feb 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button