Open official notice

Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open

Official notice

21 Dec 2016

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open. The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box. Published 2016-12-21.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

713

Recall number

Z-0769-2017

Official notice

2016-12-21

The hazard

The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

75228

Recall Numbers

Z-0769-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Updated Natus neoBLUE blanket LED Phototherapy Systems. It provides phototherapy underneath the infant and can be used in a bassinet, open bed, radiant warmer, incubator, or while holding the infant. Affected systems including: 1) Updated light boxes (serial number xxxx004283 or greater) have protection built in to turn off the device if an overheat condition is detected. 2) Updated fiberoptic blankets (Lot Nxxxx16-xx or greater) have an improved fiberoptic cable end to reduce the degradation and increase its useful life. Affected Part Numbers and Description: 006244 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad 006254 neoBLUE blanket LED Light Box 006895 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad 007299 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad EUR 007300 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad EUR 007296 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad UK 007298 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad UK 007301 neoBLUE blanket LED Phototherapy Systems with Large blanket/pad AUS 007302 neoBLUE blanket LED Phototherapy Systems with Small blanket/pad AUS

Units affected

713

Hazard

The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of the fiberoptic pads experienced by some customers on older systems. These failures on previous systems involved discoloration and melting of the fiberoptic bundle at the connector that is inserted in the neoBLUE blanket light box.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

CA · Canada

Affected area 6

CL · Chile

Affected area 7

CN · China

Affected area 8

DE · Germany

Affected area 9

DK · Denmark

Affected area 10

DO · Dominican Republic

Affected area 11

ES · Spain

Affected area 12

FI · Finland

Affected area 13

FR · France

Affected area 14

GB · United Kingdom

Affected area 15

HK · Hong Kong

Affected area 16

HU · Hungary

Affected area 17

IT · Italy

Affected area 18

LV · Latvia

Affected area 19

MY · Malaysia

Affected area 20

PL · Poland

Affected area 21

RO · Romania

Affected area 22

RU · Russia

Affected area 23

SA · Saudi Arabia

Affected area 24

SG · Singapore

Affected area 25

SI · Slovenia

Affected area 26

TR · Turkey

Affected area 27

TW · Taiwan

Affected area 28

UY · Uruguay

Affected area 29

ZA · South Africa

Company & contacts

Manufacturer

Natus Medical Incorporated

Phone

Not provided

Brand: Natus Medical. Contact: Natus Medical Incorporated.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0769-2017

Event ID

75228

Units affected

713

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Dec 2016. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button