Trumatch ct pin guide kit l- trumatch patient specific instruments are intended to be used as patient-specific surgical instrumentation to
Official notice
8 Sept 2021
An official FDA notice concerns Trumatch ct pin guide kit l- trumatch patient specific instruments are intended to be used as patient-specific surgical instrumentation to. Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment. Published 2021-09-08.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
4
Recall number
Z-2394-2021
Official notice
2021-09-08
The hazard
The hazard
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Location 4
Location 5
Location 6
Location 7
Location 8
Location 9
Location 10
Affected area 1
Affected area 2
Affected area 3
Company & contacts
Company & contacts
Brand: Depuy Orthopaedics. Contact: DePuy Orthopaedics, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.